Technical Data's

General data
  • Measuring range (dioxin & furan): 0.0001-10 ng I-TEQ / m3 (TÜV approved 0-0.2 ng I-TEQ / m3)
  • Sampling interval: from 6 hours up to 6 weeks (TÜV approved 4 weeks)
  • Flue gas temperature: up to 70°C without cooling (up to 400°C with cooling)
  • Max dust concentration in the flue gas: 50 mg/ m3
  • Flue gas velocity: from 1 up to 30 m/s
  • Operating temperature (control cabinet): 5 to +40°C (optional air conditioner for temperatures over +40°C)
  • Max relative humidity (control cabinet): 50%
  • Isokinetic control cycle: 1 sec
  • Velocity measurement accuracy: 1% of measuring range
  • Volume measurement accuracy: 1.5% of measuring range
Sampling probe
  • Length: from 350 up to 2000 mm
  • Probe shaft diameter: 60 mm
  • Free probe tip diameter: 4, 5 or 6 mm
  • Stack mounting: DN 100 flange (other flanges upon request)
  • Material: titanium (glass as an option)
Sampling unit
  • Standard unit dimensions: 650x450x250 mm (HxWxD) (or 600x600x200 mm)
  • Overbox dimensions: 1150x650x500 mm (HxWxD) (waterproof enclosure for outdoor installation)
  • Adsorbant cartridge: XAD-II
Control cabinet
  • Dimensions: 2000x800x650 mm (HxWxD)
  • Weight: approx. 250 kg
Utilities
  • Compressed air: 3 to 7 bars, dry, oil free
  • Compressed air connection: 8x1 or 6x1 mm hose
  • Cooling water: 0.5 to 5 l/min, according to fume temperature (absolutely essential for fume temperature >70°C - optional closed loop cooler if tap water not available)
  • Water connection: 1/2" hose (inlet & return)
  • Power supply: 230 VAC, 50 Hz (option 115V,50/60Hz)
  • Power consumption: approx 1.1 kW
  • Fuse: 16A
Inputs / Outputs
  • Digital outputs: status (monitoring mode, fault, error)
  • Digital inputs: furnace off, maintenance
  • Analog inputs (optional) : O2,CO2, gas velocity, static pressure etc
Disposal
  • Flue gas recycling: 8x1 mm hose
  • Condensate drain: 8x1mm hose
  • Condensate quantity: approx 3 l/day (depending on flue gas moisture content)
Options
  • Color LCD display
  • Operation using jog dial
  • Automatic multiplexing (up to 4 sampling units)
  • Internal data storage (up to 7 months of 1/2h average data)
  • TCP/IP connection
  • USB flash drive